Command Reference
CALCulate Subsystem
66
CALCulate:LLINe Subsection
Limit lines can be defined for your measurement. You can then
have the instrument compare the data to your defined limits and
indicate a pass/fail condition.
Delete All Limit Lines in Memory
:CALCulate:LLINe:ALL:DELete
Deletes all limit lines in volatile memory.
Key access:
> Limits > Delete All Limits
Define Limit Line Values
:CALCulate:LLINe[1]|2:DATA<x-axis>,<ampl>,<connected>
{,<x-axis>,<ampl>,<connected>}
:CALCulate:LLINe[1]|2:DATA?
Defines limit line values, and destroys all existing data. Up to 20
points may be defined for each limit. No units are allowed.
<x-axis> – frequency values
<ampl> – amplitude values are in the current Y-axis units. Up to
two amplitude values can be provided for each x-axis value, by
repeating <x-axis> in the data list. No unit is allowed in this
parameter.
<connected> – connected values are either 0 or 1. 1 means the
point should be connected to the previously point to define the
limit line. 0 means that it is a point of discontinuity and is not
connected to the preceding point.
Example: CALC:LLIN1:DATA 1000000000,–20,0,200000000,–30,1
Range: <x-axis>
9 kHz to 3 GHz
<ampl>
100 dBm to +30 dBm
<connected>
0 or 1
Det/
Display