Command Reference
SENSe Subsystem
150
Last Frequency Span
[:SENSe]:FREQuency:SPAN:PREVious
Set the frequency span to the previous span setting.
Key access: > Last Span
Start Frequency
[:SENSe]:FREQuency:STARt <freq>
[:SENSe]:FREQuency:STARt?
Set the start frequency.
*RST: 0 Hz
Range: –80 MHz - 3.080 GHz
Key access: > Start Freq
Stop Frequency
[:SENSe]:FREQuency:STOP <freq>
[:SENSe]:FREQuency:STOP?
Set the stop frequency.
*RST: 3.0 GHz
Key access: > Stop Freq
[:SENSe]:POWer Subsection
Input Attenuation
[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>
[:SENSe]:POWer[:RF]:ATTenuation?
Set the input attenuator. This value is set at its auto value if input
attenuation is set to auto.
*RST: 20 dB
Range: 0 to 70 dB
Key access: > Attenuation Auto Man