Command Reference
CALibration Subsystem
82
CALibration Subsystem
These commands control the self-alignment processes.
Align All Instrument Assemblies
:CALibration[:ALL]
:CALibration[:ALL]?
Performs an alignment of all the assemblies within the instrument,
except for the tracking generator.
Before executing this command, connect a cable between front
panel CAL OUT and RF IN connector.
The query performs a full alignment and returns a number
indicating the alignment is completed or not. A “0” is returned if
the alignment is completed.
Key access: > Alignments > Align > All
Coarse Adjust the Frequency Reference
:CALibration:FREQuency:REFerence:COARse <setting>
:CALibration:FREQuency:REFerence:COARse?
Allows coarse adjustment of the internal 10 MHz reference
oscillator timebase of the analyzer.
Range: Integer, 0 to 255
Key access: > Alignments > Time Base
Preset/
System
NOTE
:CALibration:ALL is required after COARse is set.