139
Command Reference
SENSe Subsystem
Set the Stop Freq
[:SENSe]:SEMask:OFFSet:LIST:FREQuency:STOP
<freq>,...[<freq>] (up to five values)
[:SENSe]:SEMask:OFFSet:LIST:FREQuency:STOP?
Specifies the inner limit (frequency closest to the carrier) for both
segments of the specified offset pair. When Start Freq (Off) is
selected, the offset pair is not measured.
Offsets that are turned off, for the currently selected offset will
return a -999.0 when queried.
*RST: Dependent upon Offset selected, please refer to the User’s Guide
for more details.
Range: 0 Hz to the Stop Freq (for that offset) minus 10 Hz
Example: SEM:OFFS:LIST:FREQ:STOP 2 MHz
Key access:
> More > Spectrum Emission Mask > Offset/Limits > Stop Freq
Set the Sweep Time
[:SENSe]:SEMask:OFFSet:LIST:SWEeptime <time> ...
[<time>] (up to five values)
[:SENSe]:SEMask:OFFSet:LIST:SWEeptime?
[:SENSe]:SEMask:OFFSet:LIST:SWEeptime:AUTO
OFF|ON|0|1,...OFF|ON|0|1 (up to five values)
[:SENSe]:SEMask:OFFSet:LIST:SWEeptime:AUTO?
Specifies the sweep time for the currently offset, and toggles this
function between Auto and Manual for each offset.
*RST: Dependent upon Offset selected, please refer to the User’s Guide
for more details.
Range: 5 ms to 4 ks
Example: SEM:OFFS:LIST:SWE:AUTO 40 ms
Key access:
>More >Spectrum Emission Mask >Offset/Limits >Sweep Time