Command Reference 
SENSe Subsystem
144
Turn Automatic Averaging On/Off
[:SENSe]:AVERage:TYPE:AUTO OFF|ON|0|1
[:SENSe]:AVERage:TYPE:AUTO?
Sets the averaging to be automatically set to the appropriate type 
for the current measurement setup. Or allows you to manually 
choose the type of averaging by 
[:SENSe]:AVERage:TYPE.
When AUTO is On:
If the Y Axis Scale is not Linear or Log, then average type is Video 
(Y Axis Scale) Averaging. If the Y Axis Scale is Linear or Log, then 
average type is Power Averaging. If the Detector is Peak, Sample, 
or
Negative Peak (not Average), then average type is Video Average.
*RST: On
Key access:  > Average Type Auto Man
Type of Averaging for Measurements
[:SENSe]:AVERage:TYPE VIDeo|RMS
[:SENSe]:AVERage:TYPE?
Successive measurements of data can be combined to average out 
measurement variations. Detector is set to average and Avg type is 
set to power (RMS) to measure RMS voltage (avg power).
*RST: VID
Key access:  > Average Type