➢ Go to “Group” with the keypad.
➢ Press OK to activate it: ON or to deactivate it: OFF.
For the groups creation and management, see page 24.
5.1.8. Select the alarm sound
“Setting > Instrument” screen is displayed.
➢ Go to “Audio” with the keypad then press OK.
➢ Press again OK
➢ Select with the keypad the Alarm 1 sound or the Alarm 2 sound then press OK.
5.1.9. Activate or deactivate Bluetooth
®
function (optional)
Bluetooth
®
allows to communicate with the Kigaz mobile application.
“Setting > Instrument” screen is displayed.
➢ Go to “Bluetooth” with the keypad.
➢ Press OK to activate it: ON or to deactivate it: OFF.
The icon appears on the instrument screen when it is activated.
When Bluetooth
®
is activated, Kigaz appears with the name “Kigaz” in the available devices list of the computer, tablet or smartphone.
5.2. Set the combustible
15 different combustibles are recorded in the analyser:
Sahara / Fos Natural gas Cannel-coal
Groninguen – Natural gas Low volatile coal
Russia – Natural gas Coke gas
Propane Biofuel 5%
LPG Wood 20%
Butane Hog fuel 21%
Light oil Pellet 8%
Heavy fuel oil
See page 21 for the measurements parameters units and values list of a factory setting.
5.2.1. Modify a combustible
Each combustible has its own coefficient allowing to calculate flue gases CO
2
, excess losses and efficiencies in the most accurate way.
The modification of these coefficients will lead some modifications in combustion analysis results. Only a qualified technician
is authorized to modify them.
It is still possible to reset the factory parameters pressing the “Factory parameters” function key or with the LIGAZ-2
software (see page 55).
It is possible to modify the name of a combustible and its type of energy (solid, liquid or gas):
“Setting > Combustibles” screen is displayed.
➢ Go to the required combustible with the up and down arrows then press OK.
➢ Press “OK” on “Name” line.
The keyboard appears.
➢ Select the letters with the direction arrows and press OK to validate.
• To go from lower-case keyboard to upper-case keyboard then to numeric keyboard: press the function key.
• To erase a letter: press the function key.
➢ Press the function key to validate the new combustible name.
➢ Press “OK” on the “Energy” line.
➢ Select the combustible energy type: solid, liquid or gas and press OK.
➢ Press the function key.
The instrument backs to the combustible list with the modifications applied.
It is also possible to modify the combustible coefficients:
“Setting > Combustibles” screen is displayed.
➢ Go to the required combustible with the up and down arrows then press OK.
➢ Go to the “Coefficient” line and press OK.
➢ Go to the coefficient to modify with the up and down arrows then press the function key.
The coefficient modification field is displayed.
➢ Modify the values with the arrows and press OK to validate.
➢ Press Esc to go back to the previous screen.
➢ Press the function key to save the new combustible.
18 Set the different parameters of the analyser