EasyManua.ls Logo

klover DIVA - Temperature Probes

klover DIVA
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10
The stove enters functioning economy upon reaching said temperature.
The stove now automatically switches off if at least one of the following conditions exist:
If it remains in Functioning Economy “ECO-H2O for a time set on Pr23 (120 minute default setting).
If it exceeds temperature differential set on Pr12 (10°C default setting) or H2O Temperature > (“SET
H2O” + Pr12)
- DHW Temperature (only on prepared models)
Press key 3 to select “SET SANI(with work power set on DHW Power) to modify DHW temperature.
Press keys 1 and 2. The display appears as in figure 9 during this operation.
S.R.L.
SET
OK
Set
°C
1
2 3 4 6
5
65°C
Wait 5 seconds after setting wanted value.
4. TEMPERATURE PROBES
The used temperature probes' features are given below.
4.1 Room/water temperature probe
The room/water probe used to measure water or room (in case of air stoves) temperature has the
following features:
IP67, NTC Thermoplastic probe: like NTC 10K ± 1% = 3435
Measuring field: -50/+120°C
Obtain a tester to check if probe is faulty, set it in ohm and measure the two probe pins (wire ends):
The probe measures about 10kohm at 25°C if it works correctly
If the probe is faulty: - resistance greater than 1Mohm (open)
- resistance to 0 ohm (in short)
The tester can also be used to check if the probe connection to the power board contacts are in short;
if 0 ohm resistance is measured by connecting the tester to the contacts, these are in short.
The lower display shows water (or room) temperature at 02°C if the room/water probe is not
connected to the power board. The display appears as in figure 10.
S.R.L.
SET
OK
Set
°C
1
2 3 4 6
5
Fig. 9
Fig. 10
02°
C

Related product manuals