Installing the Cephalometric Frankfort Positioning Laser (For Upgrade Only)
5–8 INSTALLING THE CEPHALOSTAT
Installing the Cephalometric Frankfort
Positioning Laser (For Upgrade Only)
To install the cephalometric Frankfort positioning laser, follow these steps:
1. Unscrew and remove the generator front cover.
2. Position correctly the cephalometric Frankfurt positioning laser and screw the fixing
screws.
3. Connect the cable LASER CEPH on the CJ732 board to the J28 LASER CEPH connector.
Attach the cable with a cable tie.
4. Turn on the unit. On the control panel, click to turn on the positioning laser beams.
Make sure that the cephalometric Frankfurt positioning laser beam is projected in the
middle of the ear cone. If needed, adjust the laser using the fixing screw of the laser.
2.5 mm
x2
x2