X-Ray Generator Technical Specifications
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed 01 7–5
X-Ray Generator Technical Specifications
The x-ray generator comprises the following:
• A transformer and an x-ray tube and their associated electronic components immersed in oil
• An aluminum filter, which enhances the quality of the beam and reduces the dose received by the patient
• A lead collimator, which limits the size of the beam at the image receiver unit
• A thermal cutout, which trips at an operating temperature between 63 to 70° C (± 5° C)
Measurement Conditions
kV Indirect on the peak kilovolt meter
mA Direct measurement in the circuit using an oscilloscope
Exposure time Measurement at 75% of the kV values with peak kilovolt meter
Table 7–2 Filtration of the Material in the X-ray Field
Standard Compliance
IEC 60601-1-3 Compliant
Nominal value of the inherent filtration at 70 kV 2.5 mm (0.10'') eq. Al
Nominal value of the supplementary filtration at 70 kV NA
Nominal value of the total filtration at 70 kV 2.5 mm (0.10'') eq. Al
Filtration value for the enclosure of the x-ray tube (at 100 kV) 0.2 mm (0.008")
Filtration value for the enclosure of the image receiver unit (at 100
kV)
0.2 mm (0.008")
Filtration value for the sensor case 0.8 mm (0.031") eq. Al