Post-Installation Control
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed 01 5–19
Calibrating the Cephalometic Blade Collimator
To calibrate the cephalometric blade collimator, follow these steps:
1. Access the Cephalometric Acquisition Window. In the Program pane:
• Click .
• Click
for a 30 x 30 format.
2. Launch an x-ray using the x-ray remote control. A white margin must be displayed on all
sides of the acquired image. The acquired image must be correctly centered.
3. If needed, reselect the cephalometric blade collimator parameters and relaunch the x-ray
until you obtain the correct margins (see Selecting the Cephalometric Blade Collimator
Parameters).
NOTE
This calibration is for all image formats.