Installing the Cephalostat Positioning Locker (For Upgrade Only)
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed 01 5–9
Installing the Cephalostat Positioning Locker
(For Upgrade Only)
To install the cephalostat positioning locker, follow these steps:
1. Position correctly the cephalostat positioning locker and screw the fixing screws.
2. Insert the inductive sensors tip of the cable 982/CJ822 in the hole (A). Connect the cable
connector 982/CJ822 on J3 (B).
3. From the CJ804 collimator board, solder the power supply cable 981/CJ804/J2 to the
cephalostat positioning locker, with cables:
Brown to - (C)
Red to + (D)
4. From the CJ804 collimator board, connect cable 980/CJ804/J1 to 980/CJ732/J23 on J23
(E).
IMPORTANT
When the installation of the unit is complete and the unit is turned
on, the LED on the inductive tip must lit. If the LED is not lit you
must insert more the tip to create the necessary electric contact.
CJ653
CJ699
CJ732
LF440
B
E
C
D
A