Post-Installation Control
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed 01 5–15
5. If needed, adjust the cephalostat head mechanically:
• For a horizontal adjustment, use the cephalostat arm screws (A) and (B)
• For a vertical adjustment, use the elbow tube clamps screws (C).
Adjusting the Nasion Support
To adjust the nasion support, follow these steps:
1. Lower the nasion support to a vertical position.
2. Access the Cephalometric Acquisition Window. In the Program pane:
• Click .
• Click
for a 30 x 30 format.
2.5 mm
4 mm
A
x4
x1
x1
B
B
4 mm
A
C
C