Post-Installation Control
5–16 INSTALLING THE CEPHALOSTAT
3. Launch an x-ray using the x-ray remote control. The nasion support must be perpendicular
to the ear cones.
4. In the Imaging Window, on the drawing toolbar, click . On the acquired image, draw
a perpendicular line between the center of the ear cones and the nasion support holes to
check that the nasion support is positioned vertically.
5. If needed, adjust the nasion support, using the nasion support fixing screws (A).
2 mm
A