Cephalometric Unit Technical Specifications
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed 01 7–3
Cephalometric Unit Technical Specifications
Table 8-1 describes the technical specifications for the KODAK 9000C and KODAK 9000C
3D digital imaging unit.
Table 7–1 Cephalometric Unit Technical Specification
KODAK 9000C KODAK 9000C 3D
Sensor Technology CCD
Sensor matrix 2100 x 2092 pixels
Image field 300 x 300 mm
Gray scale 16.384 - 14 bits
Magnification 1.14
Exposure time 0.1 to 3.2 sec.
Radiological exam options
• Lateral
• Frontal AP or PA
• Oblique
• Submento-vertex
• Carpus
Acquisition format size
• 18 x 18
• 18 x 24
• 24 x 24
• 24 x 30
• 30 x 30
Input voltage
• 230 / 240 V - 50/60 Hz
• 100/110/130 V - 50/60 Hz
Unit dimensions 2150mm (L) x 1595mm (D) x 2378mm (H)*
*Unit with 2225 mm (H) is also possible
Required space 2230 (L) x 2000 (D) x 2400 (H) mm
Weight of only the cephalostat
component
39 kg (85.8 lb.
)
Total weight with KODAK
9000C or KODAK 9000C 3D
199 kg (437.8 lb.
)