Post-Installation Control
5–20 INSTALLING THE CEPHALOSTAT
Checking the Image Quality
To check the image quality, follow these steps:
1. Raise the nasion support.
2. Open the head clamps.
3. Position the image quality phantom (A) at the center of the digital sensor.
4. Access the Cephalometric Acquisition Window. In the Program pane:
• Click .
• Click
for a 30 x 30 format.
5. In the Parameter pane, select:
• 64 Kv
• 15 mA
• 0.4 s
6. Launch an x-ray to acquire an image.