Post-Installation Control
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed 01 5–11
3. Launch an x-ray using the x-ray remote control. The head clamps must be correctly and
vertically positioned on the acquired x-ray image. Check that the blade is parallel to the
image border.
4. If needed, adjust the cephalostat head mechanically using the elbow tube for horizontal
inclination of the cepahlostat head. Untighten the fixing screws (A) slightly. Use the fixing
screws (B) to adjust the horizontal inclination. Tighten the fixing screws (A) when you
have reached the desired adjustment.
IMPORTANT
DO NOT adjust using the blade collimator. If necessary, adjust
using the cephalometric head.
2.5 mm
A
x4
x1
x1
B
B
4 mm
A