EasyManua.ls Logo

Konica Minolta CA-310 - Page 106

Konica Minolta CA-310
122 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
104
104
OFFSETERROR
TOOBRIGHT
NOSYNC.SIGNAL
OVER
SETMAINPROBE
PROBEERROR
DATAERROR
MEMORYERROR
FLICKERERROR
UNDER
FLICKERERROR
OVER
FLICKERERROR
VSYNCOVER
FLICKERERROR
PROBETYPE
Error Message
Cause: (Description) Corrective Action
Zero calibration has not been performed
correctly.
(Zero calibration was performed with insuf-
ficient blocking of entry of light.)
Pressure is given to purobe.
Perform zero calibration again. (Page 34)
(
Even if the error message is currently displayed,
measurement will start if the measuring probe’s re-
ceptor is exposed to light.
)
Don
t give the pressure.
Zero calibration is being performed with
insufficient blocking of entry of light.
Block the light completely for all the
measuring probes, and when “DARKEN
PROBE PUSH 0–CAL KEY” appears
press the 0–CAL key again. (Page 34)
Although EXT SYNC mode is selected, the verti-
cal synchronizing signal used for the display is not
input correctly to the terminal on the instrument.
The vertical synchronizing signal used for the dis-
play is outside the range of 40 to 200 Hz.
Input the vertical synchronizing signal correctly.
(When the vertical synchronizing signal is out-
side the range of 40 to 200 Hz/page 28)
Change SYNC mode to NTSC, PAL, UNIV or
INT mode and start measurement.
The measured value is exceeding the
instrument’s measurement range.
The measured value is above 100,000% in
analyzer mode. (Display range over)
Measurement must be performed within
the measuring range.
The measuring probe is not connected to the
probe connector [P1] properly.
Connect the probe to the probe connec-
tor [P1] properly.
(Before connecting/disconnecting the
measuring probe, make sure that the
POWER switch is set to OFF.)
A measuring probe was connected or dis-
connected while the POWER switch was
ON.
Set the POWER switch to OFF first, con-
nect the measuring probe, then set the
POWER switch to ON.
(Before connecting/disconnecting the
measuring probe, make sure that the
POWER switch is set to OFF.)
Measurement is not possible since the mea-
suring circuit is not functioning correctly.
Set the POWER switch to OFF.
If this error still appears even if the
POWER switch is set to ON, the instru-
ment has broken down.
Contact a Konica Minolta authorized ser-
vice facility.
The instrument’s memory is abnormal.
If the message disappears when the probe
is changed, the probe's memory is abnormal.
If the message keeps appearing when the probe is
changed, the main body's memory is abnormal.
Set the POWER switch to OFF.
If this error still appears even if the
POWER switch is set to ON, the instru-
ment has broken down.
Contact a Konica Minolta authorized ser-
vice facility.
Lv is below the instrument’s flicker mea-
suring range.
Measurement must be performed within
the measuring range.
Flicker value
(contrast method) has ex-
ceeded 100.0%.
Measurement must be performed within
the measuring range.
VSYNC is exceeding 130 Hz in flicker
mode.
In the case of flicker mode, VSYNC of
40 to 130 Hz must be input.
*3
*4
*3
*4
*3
*5
*3
*4
A measured probe not for LCD flicker mea-
suring is selected in Flicker Mode.
In Flicker Mode, select a measured probe
for LCD flicker measuring.
Contact a Konica Minolta
-
authorized
service facility.
Contact a Konica Minolta
-
authorized
service facility.

Table of Contents

Related product manuals