14 Troubleshooting
14.1 Fault table
Further fault Cause and elimination
When the electronic unit is
switched on, the following
alarms are displayed:
o 8.07Alarm. Max.
flowA exceeded.
Check pick up!
o 8.10Alarm. Analog
output1 or 2 scaling
exceeded!
o 8.11Alarm. Change
of directionA
exceeded, check 30/
s signals!
o 8.12 Alarm. Change
of directionB
exceeded, check 30/
s signals!
o 8.21 Alarm. Max.
flowB exceeded.
Check pick up!
o Use a power pack 24VDC 15W or insert a debounced switch between the electronic unit and
power pack.
o Shield the lines to the pick ups and terminate the shield to Gnd (chassis) or ground.
POWER FAIL no
counting
Falls below supply voltage 17.5V
Keyboard background
illumination flashes.
There is an input error.
o + Press simultaneously.
-> The existing errors are displayed.
o Eliminate errors.
Overflow of the total value After an overflow of the total value the electronic unit displays the following:
o For 3 decimal places:
±1999999,999
o For 1 decimal place:
±199999999,9
o Specify another unit for total, e.g. m
3
instead of l. After the modification, the total value is still avail-
able after the overflow.
o Reduce the number of decimal places.
Operating instructions
OIE 12en-GB Edition 2020-03
41