EasyManua.ls Logo

L3 comminications Trilogy ESI-2000 - Figure 2-9: Battery Calibration Menu

L3 comminications Trilogy ESI-2000
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ESI-2000
Pilot’s Guide 2-7
Basic Operation
Take the following into consideration before performing the
battery capacity meter calibration:
The battery SOC must be ≥ 90% before the calibration of
the capacity meter can be accomplished. Refer to charging
instructions in step 6.
The ambient temperature of the ESI-2000 must remain
between 10°C and 25°C (50°F and 77°F) for at least 2 hrs.
The battery must not have been in operation (charging or
discharging) for at least 2 hrs.
Apply power to the ESI-2000.
Remove external power from the ESI-2000 by placing the aircraft
battery switch(s) OFF.
The ESI-2000 automatically switches over to battery power. Press
any button as stated by the message.
Press the M (menu) button and cycle through the menu options
until Batt Cal (see Figure 2-9) is located.
Press and hold the + (Hold) button until the Battery Calibration
screen is shown with the following message:
1.
2.
3.
4.
5.
6.
Figure 2-9: Battery Calibration Menu
BATTERY CAPACITY METER CALIBRATION PROCEDURE
(Continued)
Battery Calibration In Progress...
Auto-Off When Done
The “CAL DUE” battery indicator continues to be observed
during normal operation if the calibration procedure is aborted.
Press Any Key to Abort

Table of Contents

Related product manuals