Echo Liquid Handler Software 5-5
Create a new 384:384 protocol definition
Create a new 384:384 protocol definition using the ADD NEW PROTOCOL
WIZARD.
Note: If the new liquid transfer protocol requires a new destination
microplate definition, create it before running the A
DD NEW PROTOCOL
WIZARD. See “Manage labware specifications” on page 5-24.
1. Click the ADD button to start the ADD NEW PROTOCOL WIZARD. Enter the
following information:
Protocol Name: Enter a name for the new protocol. It must not
already be in use.
Source Plate Type: Select a source plate type from the first drop-
down list.
Note: Only Echo qualified source microplates are listed in the Source
Plate Type drop-down list. Contact Labcyte to add more source
plates. See “Contact information” on page 7-1.
Destination Plate Type: Select a destination plate type from the
second drop down list. The destination plate definition must already
exist in the Echo database.
Figure 5.3 Add new protocol wizard, page 1 of 2