Echo Liquid Handler Software 5-55
Save Transfer data to disk: Normally, the
data collected during a source plate transfer is
deleted when the transfer is complete. If the
S
AVE TRANSFER DATA TO DISK checkbox is
selected, the transfer data will be stored on the client PC disk. There is no
reason to save transfer data during normal operation. The data file is stored
to <InstallDir>\Labcyte\Medman\Data. Transfer data can be useful to
Labcyte personnel during diagnostic troubleshooting.
Default: liquid transfer data is not saved to the Echo controller.
Ignore Source Plate Sensor Check: Selecting
the IGNORE SOURCE PLATE SENSOR CHECK
checkbox bypasses the source plate present
sensor. This may be useful when an individual source plate does not fit
correctly. A run protocol procedure can execute without triggering a “no
source plate loaded” error message.
Default: The source plate presence detect sensor is not ignored.
Ignore Destination Plate Sensor Check:
Selecting the IGNORE DESTINATION PLATE SENSOR
CHECK checkbox bypasses the destination plate
present sensor. This may be useful when an
individual destination plate does not fit correctly, or droplets are being
transferred to a microplate lid or other substrate that does not fit like a
microplate. A run protocol procedure can execute without triggering a “no
destination plate loaded” error message.
Default: The destination plate presence detect sensor is not ignored.
5.5.2 Advanced solvent concentration (DMSO
only)
If the DMSO concentration in a source plate is known, accuracy can be
improved by specifying it prior to performing liquid transfers.
Note: The Solvent Concentration function assumes the solvent is
DMSO, therefore this function is not relevant for aqueous solutions.
The DMSO concentration information entered will be used with each
source microplate until one of the following occurs:
The USE PREDEFINED checkbox is cleared, or
New DMSO concentration information is entered, or
The Echo software is restarted.