EasyManuals Logo

LeCroy 9410 User Manual

LeCroy 9410
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #48 background imageLoading...
Page #48 background image
System Commands
I
I
MISCELLANEOUS
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
Response Format
EXAMPLE (GPIB)
RELATED COMMANDS
AUTO_CALIBRATE, ACAL
Command/Query
The AUTO CALIBRATE command is used to enable or disable
automatic calibration of the instrument. At power-up, auto-cali-
bration is turned ON, i.e. all input channels are periodically
calibrated for the current input amplifier and time-base settings.
The automatic calibration may be disabled by issuing the com-
mand ACAL OFF. Whenever it is convenient, a *CAL? query
may be issued to fully calibrate the oscilloscope. When the oscillo-
scope is returned to local control, the periodic calibrations will be
resumed.
The response to the AUTO CALIBRATE? query indicates
whether auto-calibration is enabled.
Auto_CALibrate <state>
<state> := {ON, OFF}
Auto_CALibrate?
Auto_CALibrate <state>
The following instruction disables auto-calibration.
CMD$="ACAL OFF": CALL IBWRT(SCOPE%,CMD$)
*CAL?
I
I
I
I
I
I
I
I
I
I
I
I
I
|
42

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LeCroy 9410 and is the answer not in the manual?

LeCroy 9410 Specifications

General IconGeneral
BrandLeCroy
Model9410
CategoryTest Equipment
LanguageEnglish

Related product manuals