WAVERUNNER XI SERIES
WARNING
This is a Class A product. In a domestic environment this product may cause radio interference, in which case the
user may be required to take appropriate measures.
Electromagnetic Immunity:
EN 61000-4-2/A2:2001* Electrostatic Discharge
(4 kV contact, 8 kV air, 4 kV vertical/horizontal coupling planes)
EN 61000-4-3/A1:2003* RF Radiated Electromagnetic Field
(3 V/m, 80-1000 MHz)
EN 61000-4-4:2004* Electrical Fast Transient/Burst
(1 kV AC Mains, 0.5 kV I/O signal/control)
EN 61000-4-5/A1:2001* Surges
(1 kV AC Mains, 0.5 kV I/O signal/control)
EN 61000-4-6/A1:2001* RF Conducted Electromagnetic Field (1 kV /
0.5 kV common mode / differential mode - AC Mains)
EN 61000-4-11:2004
Mains Dips and Interruptions (1 cycle voltage
dip, 100% short interruption)
* Meets Performance Criteria "B" limits during the disturbance; product undergoes a temporary degradation or
loss of function of performance which is self recoverable.
Meets Performance Criteria "C" limits during the disturbance; product undergoes a temporary degradation or
loss of function of performance which requires operator intervention or system reset.
Low Voltage Directive:
EN 61010-1:2001
Safety requirements for electrical equipment for measurement,
control, and laboratory use.
The oscilloscope has been qualified to the following EN 61010-1
limits:
Installation Categories II (Mains Supply Connector) & I (Measuring
Terminals)
Pollution Degree 2 (Normally only dry non-conductive pollution
occurs. Occasionally a temporary conductivity caused by
condensation must be expected.)
Protection Class I (Provided with terminal for protective ground)
UL and cUL Certifications:
UL Standard: UL 61010-1 2
nd
Edition
Canadian Standard: CSA-C22.2 No. 61010-1-04
Warranty
The instrument is warranted for normal use and operation, within specifications, for a period of three years from
shipment. LeCroy will either repair or, at our option, replace any product returned to one of our authorized service
centers within this period. However, in order to do this we must first examine the product and find that the defect
is due to workmanship or materials and not due to misuse, neglect, accident, or abnormal conditions or operation.
LeCroy shall not be responsible for any defect, damage, or failure caused by any of the following: a) attempted
repairs or installations by personnel other than LeCroy representatives, or b) improper connection to incompatible
22 WRXi-OM-E Rev C

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LeCroy WaveRunner XI Series and is the answer not in the manual?

LeCroy WaveRunner XI Series Specifications

General IconGeneral
BrandLeCroy
ModelWaveRunner XI Series
CategoryTest Equipment
LanguageEnglish

Related product manuals