NOVOLAS Basic AT Version 2.1
09/2010
44 Operation
Register
Index
Name Function Units
Value range
512 INTERLOCK: Common error in
interlock testing.
Restart. Replace module if necessary
(bei Leister service)
1024 LASERVOLTMAX: Laser voltage
exceeded maximum.
Restart. Replace module if necessary
(bei Leister service)
2048 LASERVOLTMIN: Laser voltage
dropped below minimum.
Restart. Replace module if necessary
(bei Leister service)
4096 SAFETY: Common error in safety
module.
Restart. Replace module if necessary
(bei Leister Service)
8192 CUSTOMERIO LINK: Ethernet
communication with CustomerIO
failed
16384
OPTICINTERFACE_LINK: Ethernet
communication with OpticInterface
(pyrometer) failed
32786
ILOPENEDWHENLASERON: The
interlock was opened when the laser
was on
Register
Index
Name Function Units
Value range
0 Off: Current source is off.
1 ReadyInactive: Current source
delivers no current.
2 ReadyTestCurrent: Current source is
activated and delivers the test
current.
3 ReadyActive: Current source is
activated and delivers the
programmed current.
121 Status Actual status value
4 Error: Current source is in error state.
9910 Not used
The function of the values in register 9911 depend on the laser module that is used
0 NO ERR: no error 9911 Interlock: Air
Cooled Laser
Returns a value
describing the
1 TEMP_1: Error Temperature 1