23
Medical electrical equipment or medical electrical system should not be used in extremely short
distances within the bed! In the case it is necessary to use medical electrical equipment in this
extremely short distances within the bed observe this configuration regularly to verify the normal
functions of the bed in this configuration.
Consult Linet
®
or local dealer before using medical electrical equipment where influence on the bed can be
strong.
NOTE AAMI/ANSI/IEC certified medical electrical equipment used in ICU is not restricted.
NOTE It does not concern fixed installations of hospital communication systems and nets (DECT,
pager, WiFi) and other mobile devices which are certified according to local telecommunication legislation.
The birthing bed must be installed and put into operation in compliance with information related to EMC
provided in the accompanying documentation. Portable and mobile HF communication devices may
influence the electrical medical device.
Before plugging check the power cable for mechanical damage.
Before plugging in check the voltage of the network for conformity with information on the
type label of the bed. A plugged-in bed is indicated by a green LED indicator on the side control
panel of the bed.
Power cable of the bed must not be connected through an extension cord. Only connection
possible is directly in the power socket.
Avoid hanging the power cable over moving parts of the bed. When back up batteries are in use,
the cable may be damaged!
No input cable can lead over the bed and nor over the patient.
Within the radius of 1.50 m, within the patient space, only medical equipment meeting standard
AAMI/ANSI ES 60601-1 may be used.
Birthing bed is not intended for use in areas enriched with oxygen, i.e. more than 25% or partial
oxygen pressure 27,5 kPa.
Manual and declaration of the manufacturer - electromagnetic radiation
The birthing bed AVE 2 is intended for use in electromagnetic environment. The customer or user of the birthing
bed must ensure that the bed is used in such environment.
Radiation Test
Conformity
Electromagnetic Environment
High-frequency radiation CISPR 11
Group 1
AVE 2 utilises high-frequency
energy for its internal function
only. The high-frequency
radiations are very low and
unlikely to cause any interference
to nearby electronic devices.
High-frequency radiation CISPR 11
Class B
AVE 2 is suitable for all
institutions, including households
and objects directly connected to
the public low-voltage mains
supplying residential buildings.
Harmonic radiations IEC 61000-3-2
Class A
Fluctuating voltage/Flashing
radiation CE 61000-3-3
Satisfactory