SCP System • Error messages and error diagnosis
26 CP_SEM_60-00-00.002
4.1.2 Error codes
The following error messages are only shown on the 7-segment display “Actual value” (RPM).
Error Description/Possible causes Corrective measures
(please observe sequence)
RAM defective Replace CPU board
see page 32
EPROM defective, checksum incorrect Replace EPROM or
–Replace CPU board
see page 31
see page 32
EEPROM defective Replace CPU board
see page 32
Processor error Replace CPU board
see page 32
Display error, either Display board or
CPU board defective
First replace display board; if the error
display persists
–Replace CPU board
see page 33
see page 32
Processor error Replace CPU board
see page 32
Difference set value-actual value of pump
more than 250 RPM,
drive unit error
–Replace motor control board
see page 40
A/D converter: reference voltage outside
tolerance limits
–Replace CPU board
see page 32
Shut-off path test (on powering up)
defective
–Replace CPU board
see page 32
Difference set value-actual value of pump
more than 1000 RPM, drive unit or
CPU board error
–Replace motor control board; if the
error display persists
–Replace CPU board
see page 40
see page 32
E01
E02
E04
E06
E08
E09
E62
E64
E71
E76