xMAP Technology The System
9
• Reporter control microspheres (CON2)
• Luminex xMAP Sheath Fluid
xPONENT 3.1 • Calibration Kit (CAL)
• Verification Kit (VER)
• Luminex xMAP Sheath Fluid
Required
Laboratory
Reagents
• Household bleach
• 70% isopropanol or 70% ethanol
• Mild detergent
• Distilled water
Luminex Software Luminex software provides complete control of the system and
performs data analysis. Your Luminex 200 system is preloaded with
the Luminex software. However, we supply a software DVD should
you need to reinstall the software.
This software requires a dedicated system. Unauthorized additional
software is prohibited and may result in improper operation of the
system.
Luminex 200
Performance
Specifications
Speed • USB communications link for fast data transfer
• Automatic transfer of assay templates and new reagent
information into the system via a large capacity read/write DVD
• Installation: < 4 hours
• System calibration: < 10 minutes
• System controls: < 10 minutes
• Barcode reader entry of sample IDs
• Automatic post-analysis
• Analyze one 96-well plate/hour depending on manufacturer’s kit
• Up to 100 xMAP microsphere sets per sample
• System warmup: 30 minutes. Systems that remain inactive for at
least four hours will require a warm-up to restart the lasers. After
acquiring sample, running system calibrators, running system
controls, and warming up the instrument, the system resets the
four-hour internal clock.