xMAP Technology The System
13
UL listed (or equivalent), CSA certified, and CE marked when used
internationally.
Surge Protector If you do not use a UPS, use a surge protector. Choose a protector
that meets your needs. Factors to consider include electrical
environment, endurance, suppressed voltage rating, and method of
protection. It should have six outlets, rated at least 1500 Watts, and
be UL listed, CSA certified, CE marked for nondomestic use when
used internationally.
Printer Printer, HP LaserJet 2300 or available equivalent
Barcode Labels Use the Code 128 barcode label type when scanning barcode labels
into the system as patient identities.
Vortex VWR product number 58816-121: Speed range 0-3200 rpm or
equivalent
Bath Sonicator Cole-Parmer® product number 08849-00: Operating frequency 55
kHz or equivalent
System Overview The system consists of three subsystems: electronic, fluidic, and
optical. The following section describes the user-accessible
components of each subsystem.
Electronics
Power Input
Module
The power input modules contain the on/off switch and fuses.
Communications
Ports (DB9-PIN)
The communications port connects the Luminex 200 analyzer or the
Luminex XYP instrument to the computer, and the Luminex SD
system to the Luminex 200 analyzer.
Luminex 200
Analyzer
Ventilation Filter
Located on the bottom of the Luminex 200 analyzer, the filter must
be checked and cleaned as necessary. For proper ventilation, do not
obstruct the area below and allow at least two inches (5 cm) of
clearance around the Luminex 200 analyzer.
Luminex XYP
Instrument
Ventilation Filter
The XYP instrument ventilation filter cleans the air that cools the
internal parts of the Luminex XYP instrument. See Figure 1.

Table of Contents

Other manuals for Luminex 200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Luminex 200 and is the answer not in the manual?

Luminex 200 Specifications

General IconGeneral
Power Requirements100-240 VAC, 50/60 Hz
Dynamic Range3.5 logs
Number of AnalytesUp to 100
SoftwarexPONENT
ApplicationImmunoassays, Nucleic Acid Detection, Enzyme Assays
TechnologyxMAP (Multi-Analyte Profiling)
Bead TypesColor-coded microspheres
Lasers2 (532 nm and 635 nm)
Assay TypesImmunoassays, nucleic acid assays, enzyme assays
Sample Volume12.5 - 50 µL
PMT DetectorsYes