12
Luminex XYP
Instrument General
Ambient temperature: 15°C to 30°C (59°F to 86°F)
Humidity: 20% to 80%, noncondensing
Altitude: operation up to 2400 m (7874 ft) above mean sea level
Physical dimensions: 44 cm (17.25 inches) W x 60 cm (23.5
inches) D x 8 cm (3 inches) H
Weight: 15 kg (33 lbs.)
UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
Heater operating range: 35°C to 55°C (95°F to 131°F) with
tolerance 0°C to +2°C
Input voltage range: 100-240 V~ ± 10%, 1.8 Amps, 47-63 Hz
AC inlet fuse: 3 A, 250 V~, fast acting
Luminex SD System
General
Ambient temperature: 15°C to 30°C (59° to 86°F)
Humidity: 20% to 80%, noncondensing
Altitude: designed to operate at up to 2400m (7874 feet) above
mean sea level
Physical dimensions: 20 cm (8 inches) W x 30 cm (11.75 inches)
D x 24.75 cm (9.75 inches) H
Weight: 9 kg (20 lbs)
UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
Input voltage range: 100-240 V~ ± 10%, 0.4 Amps, 47-63 Hz
AC inlet fuse: 2 Amp, 250 V~, time lag
PC Specifications For updated information regarding the PC, notebook, or operating
system, go to http://www.luminexcorp.com, then click on the
Support link to open the FAQ list.
Recommended Additional
Equipment
Uninterruptible
Power Supply
(UPS)
Luminex highly recommends using an uninterruptible power supply
(UPS) to protect your system from power outages. Choose one that
can provide 1050 Watts for at least 45 minutes. The UPS should be

Table of Contents

Other manuals for Luminex 200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Luminex 200 and is the answer not in the manual?

Luminex 200 Specifications

General IconGeneral
Power Requirements100-240 VAC, 50/60 Hz
Dynamic Range3.5 logs
Number of AnalytesUp to 100
SoftwarexPONENT
ApplicationImmunoassays, Nucleic Acid Detection, Enzyme Assays
TechnologyxMAP (Multi-Analyte Profiling)
Bead TypesColor-coded microspheres
Lasers2 (532 nm and 635 nm)
Assay TypesImmunoassays, nucleic acid assays, enzyme assays
Sample Volume12.5 - 50 µL
PMT DetectorsYes