Luminex 200 System User Manual
22
To clean the sample probe:
1. Remove the sample probe as follows: Unsnap the light housing
located above the probe. Then, unscrew the Cheminert fitting on
top of the probe completely. Next, gently grasp the probe and
push up. Remove the probe out of the top of the sample arm.
2. Remove the sample probe and sonicate the narrow end for 2-3
minutes. Keep the larger end out of the sonication fluid.
3. Using a syringe, flush the sample probe with distilled water from
the narrow end out through the larger end.
4. Replace the sample probe and readjust the height for the plates
you are using.
5. Run 3 backflushes, 3 drains, 2 alcohol flushes, and 3 washes
with distilled water.
Every Six Months
Luminex 200 Analyzer Air
Intake Filter
To replace the analyzer air intake filter:
1. Disconnect the Luminex 200 analyzer from AC power by turning
off the power switch on the rear of the analyzer, then unplugging
the power cord from the wall source.
2. On the back of the Luminex 200 analyzer, in the upper left
corner, remove the screw at the top of the panel and open the
panel door.
3. Grasp the tubing and pull the filter 3 to 4 inches from the unit.
See Figure 4.
Caution: The Luminex analyzer sample probe should slide up eas-
ily while removing it from the sample arm. If you feel
resistance, do not force the probe up. Contact Luminex Technical
Support.

Table of Contents

Other manuals for Luminex 200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Luminex 200 and is the answer not in the manual?

Luminex 200 Specifications

General IconGeneral
Power Requirements100-240 VAC, 50/60 Hz
Dynamic Range3.5 logs
Number of AnalytesUp to 100
SoftwarexPONENT
ApplicationImmunoassays, Nucleic Acid Detection, Enzyme Assays
TechnologyxMAP (Multi-Analyte Profiling)
Bead TypesColor-coded microspheres
Lasers2 (532 nm and 635 nm)
Assay TypesImmunoassays, nucleic acid assays, enzyme assays
Sample Volume12.5 - 50 µL
PMT DetectorsYes