Electronic Fluxmeter EF 14
BA
no.: 100514EBA02
December 9
, 2020
Page
41
If the peak function is enabled, ?MEAS returns the peak value.
NMEAS Number of values for the cyclic output.
Write: #NMEAS n n = 0: cyclic output runs unlimited
n > 0: cyclic output stops automatically after n
values (max. 65535).
Answer: NMEAS n
Read: ?NMEAS
Answer: NMEAS n
MULTI Start or stop cyclic output.
Write: #MULTI 0 stop cyclic output
Answer: MULTI 0
Write: #MULTI 1 start cyclic output with counter
Answer: MULTI 1
xxxx: value unit
xxxx is a four digits hexadecimal number that gives the serial
number of the value. Counting starts with 0000.
Write: #MULTI 2 start cyclic output without counter
Answer: MULTI 2
The EF 14 starts to send values in the following format:
value unit
Read: ?MULTI
Answer: MULTI n n = 0, 1, 2
SPEED Enable or disable fast data mode (see chapter 5.4)
Write: #SPEED x
Answer: SPEED x
Parameter x:
27 to 254 enable fast data mode with a distance of x · 0.732 ms between 2
readings
255 disable fast data mode
As soon as the fast data mode is enabled, the instrument starts to send data.
Only numerical values are sent, without an unit. If required, the unit can be
requested using ?UNIT or ?MEAS before.
During the fast data mode the EF 14 accepts any other commands despite of
#SPEED and #RESET.