55
Then check the measuring coil with an Ohmmeter for wire failure.
3. Fault: The instrument is showing excessive drift that cannot be compensated using the
drift function.
Measure: Make sure that the input connector of the EF 14 and the plug of the coil have
the same temperature. Check if the coil wire is insulated from ground, shield or
other voltage potentials. Check the measuring coil with an Ohmmeter for wire
failure.
8.5 Taking Out of Operation
Switch the device off and store it in a safe place under suitable environmental conditions to
allow future reuse. It is advisable to register the date of the last operation.
Important!
To prevent any possible environmental pollution or violation of environmen-
tal regulations, only a specialized company should carry out disposal.