Electronic Fluxmeter EF 14
6.3.6 Interface Commands
LOCAL Enable the keypad again after it has been disabled by a write command
Write: #LOCAL without parameter: keypad enabled.
Answer: LOCAL 1
Read: ?LOCAL
Answer: LOCAL 0 Remote mode (keypad disabled)
LOCAL 1 Local mode (keypad enabled)
The Remote mode is automatically enabled when the EF 14 receives any
write (#) command.
TERM Terminator that the EF 14 uses to terminate an answer
Write: #TERM 0 Terminator Carriage Return
#TERM 1 Terminator Carriage Return and Line Feed
Answer: TERM x x = 0, 1
Read: ?TERM
Answer: TERM x x = 0, 1
CLR Sets all user settings, which remain stored when the EF 14 is switched off,
back to default values. Write only (#CLR), no parameter
The default settings are:
Coil: Field Coil
Unit: Vs
ID: 0
k: 1
R: 0
Range: 10
-4
Max./Min. off
Peak off
Limit off
Limit-Modus absolute
Upper limit 0
Lower limit 0
AC/DC DC
Var. off
Var.-Wert 1
Contrast mid setting
Drift mid setting
Memory erased
Coil data memory erased
NMEAS 0
Baud rate 9600