Electronic Fluxmeter EF 14
BA
–
no.: 100514EBA02
December 9
, 2020
Page
48
Terminator Carriage Return and Line Feed
STAT Instrument status, Read only (?STAT), the following answers can appear:
STAT WAIT The instrument is not yet ready for measurement.
STAT DRIFT Drift function is running.
STAT INACT The instrument cannot send a measurement, e.g. if a
menu is activated in manual operation.
STAT OK The instrument is ready for measurement.
BAUD Baud rate, write only (#BAUD x).
The communication is interrupted after changing the baud rate and must be
re-initialized from the computer. Therefore the command does not return an
answer.
UPD Update, write only (#UPD).
Forces the display content to be redrawn.