Electronic Fluxmeter EF 14
Answer: PEAK n n = 0, 1
MPEAK Request of the peak value
Read: ?MPEAK
Answer: MPEAK value unit
The command returns an answer only if the Peak function is enabled.
As the captured peak value is shown as the main value, it can also be
requested using ?MEAS.
MVALLEY Request of the valley value
Read: ?MVALLEY
Answer: MVALLEY value unit
The command returns an answer only if the Peak function is enabled.
6.3.3 Instrument Functions
RESET Reset function, write only, no parameter
Write: #RESET
Answer: RESET
DRIFT Automatic drift correction, write only
Write: #DRIFT 0 ends the automatic drift correction. The established
correction value is applied (like Enter in manual
operation).
#DRIFT 1 starts the automatic drift correction
Answer: DRIFT 1 drift correction has started
If the drift correction was started by #DRIFT 1 the end is reported with the
answer DRIFT END, regardless whether the function finished automatically
or by #DRIFT 0.
When the drift correction is active, only the #DRIFT 0 write command is
accepted.
To find out, if the drift correction is still active, the ?STAT commend can be
used.
The DRIFT command is not available in AC mode.
UNIT Unit of the measured values, available parameters (0...27) depend on coil
type.
Write: UNIT x (x = 0...27)
0 Vs magnetic flux in Volt · Seconds for all coils
2 Vs cm magnetic dipole moment for moment coil