Electronic Fluxmeter EF 14
LIMIT enable and disable limit function
Write: #LIMIT 0 disables the function
#LIMIT 1 enables the function
Answer: LIMIT x (x = 0, 1)
Read: ?LIMIT
Answer: LIMIT x (x = 0, 1)
LMODE mode of the limit function
Write: #LMODE 0 absolute
#LMODE 1 +/-
Answer: LMODE x (x = 0, 1)
Read: ?LMODE
Answer: LMODE x (x = 0, 1)
LVAL Write and Request limit values, parameter x: upper limit,
y: lower limit
Write: #LVAL x, y
Answer: LVAL x unit, y unit
If a value for the upper limit is sent that is lower than the value for the lower
limit, the EF 14 exchanges the values.
The unit needs not to be sent in the write command. The EF 14 expects the
values in the actual basic unit without prefix (e.g. in T, not in mT; to set 1 mT
send 0.001).
Read: ?LVAL
Answer: LVAL x unit, y unit
LSTAT Limit status, for request only
Read: ?LSTAT
Answer: LSTAT LOW
LSTAT OK
LSTAT HIGH
The command returns only an answer if the Limit function is switched on.
6.3.5 Coil Data Commands
COIL Coil type, possible parameters: 0 … 6
0 for Field Coil
1 for Surrounding Coil
2 for Moment Coil
3 for Potential Coil