Electronic Fluxmeter EF 14
4 for Pole Coil
5 for Flux Etalon
6 for Integrator Mode
Answer: COIL x (x = 0...6)
Read: ?COIL
Answer: COIL x (x = 0...6)
For compatibility with the EF 5, the EF 14 accepts instead of the command
COIL as well the command DEVICE.
CONST Coil constant
Write: #CONST x x: numerical value 0.000001 ≤ x ≤ 90000000
Depending on the coil type, the following unit for the coil constant is
assumed:
cm² for Field Coil (area turns, winding area)
no unit for Surrounding Coil (number of turns)
cm for Moment Coil (measuring constant)
kA/Vs for Potential Coil (measuring constant)
cm² for Pole Coil (area turns, winding area)
The parameter must be converted into the required unit, if necessary. It is not
necessary to send the unit to the EF 14.
The coil types 5 and 6 do not require a constant.
Answer: CONST x
Read: ?CONST
Answer: CONST x
OHM Coil resistance in Ohm, parameter: number ≥ 0, max. 90000000
Write: #OHM x
Answer: OHM x
Read: ?OHM
Answer: OHM x
ID Coil identification number, Parameter: number ≥ 0, max. 9999
Write: #ID x
Answer: ID x
Read: ?ID
Answer: ID x