Electronic Fluxmeter EF 14
BA
no.: 100514EBA02
December 9
, 2020
Page
9
1.1.5 Safety Checks
All safety checks as described here must be carried out regularly and carefully. Please use the
checklistGeneral Check for safety checks
Check Intervals:
Once a year and
after every repair
Contents of check:
specified condition specified function
Table 1: Checklist for safety check
General Check and Functional Tests
General Check:
Switch off the device and disconnect it from mains power supply.
Visual control check of instrument casing
undamaged
Visual control check of connection cables
undamaged
Functional test:
Before functional test the device must be connected to mains power supply.
EF 14
switch on
with power switch
Display
Display lights up
EF 14
switch off
with "Power Off"
-
key
Display
extinguishes
Check Date:________________ Operator (Signature): ________________________
Important!
We recommend having the instrument and, if appropriate, the measuring
coils periodically recalibrated to ensure traceability of the measuring results.
The suitable recalibration interval depends on the operating conditions and of the quality
management stipulations of the owner. Intervals of one to two years are usual, depending
whether the instrument is continuously used in rough production environment or from time to
time under laboratory conditions.