EF 14 to low sensitivity data acquisition systems. The voltage range for a valid output is limited
to about ±4.9 V. The auxiliary output can have a considerable offset voltage of several
millivolts. Therefore it is recommended to measure the offset voltage during a reset and
subtract it from the voltage that is obtained during the measurement.
Auxiliary analog output AC-RMS, pin 14: This output supplies a voltage that is a function the
RMS value in AC mode. The RMS value is proportional to u
out
- 2.5 V. The measurements are
valid in an output voltage range between approximately 2.525 V and 4 V.