7.2.8 Editing Frequency Programs of the Application List in the Testing Mode
If you want to reprocess the frequency programs, that are stored in the application list, in
the Testing Mode, proceed as follows:
Option 1: Selection via the Navigation Bar
1. Tap the button on the far left in the navigation bar. All frequency programs of the
highest hierarchical level are listed.
2. Tap the green button labelled “Appl.”.
3. Tap the button “Down” in the navigation bar. Now the first frequency program of the
Application-screen is loaded. Or tap the button “…”. Now all frequency programs of
the application list are listed. Tap on a program to load it.
By using the buttons “Prev” and “Next” of the navigation bar, you can load the previous or
next frequency program of the application list.
Option 2: Call the Testing Mode during the Running Application
1. Select the desired frequency program in the application screen by means of the
rotary encoder.
2. Start the application by pressing the “On” key of the keypad. The currently selected
frequency program is executed.
3. Switch to the Testing Mode by clicking on the tab “Testing”.
4. The desired frequency program is now loaded in the Testing Mode and executed.
Please note that it is not automatically ended or switched to the next frequency
program when the execution time is over. This must be done manually. Now the
application list with all programs is loaded to the navigation bar.
After a frequency program was loaded from the application list with one of the two options
described, you can modify it as needed. Please note that any changes that you make in a
frequency program will not be applied to the next frequency program in the application list.
In order to make it immediately evident that frequency programs from the application list are
loaded in the Testing Mode, the corresponding buttons in the navigation bar are shown in
green.
Applying the Modifications
Press the button “Add program” if you want to re-apply the changes that have been made in
the Testing Mode to the Application Mode. The dialogue “Add … to Application” now offers
the option to replace an existing program.
If you have determined that a frequency program is no longer needed, you can tap on the
button “Remove Program” which removes the program from the application list.
Please note: the same signals are generated if the frequency generators are activated (key
“On” is lit, 3 little lamps on the touchscreen flash green) and the same settings in the Testing
Mode and Application Mode were selected.