Rad-5/5v Chapter 4: Operation
www.masimo.com 28 Masimo
Advanced Operation
The following information is designed for advanced users of the Rad-5/5v.
Default Settings
The Rad-5/5v store two types of default values: those that the device automatically resets to
after a power cycle, and those that can be changed by the user and are saved after a power
cycle.
The following table outlines the default values that the Rad-5/5v revert to after a power
cycle:
Option
Default Setting
Brightness display
Set to pre-power down setting
Pulse tone volume
Set to pre-power down setting
The following table outlines the default values that the Rad-5 reverts to after a power cycle:
Option
Default Setting*
SpO
2
high alarm limit
Set to Off
SpO
2
low alarm limit
Set to 90%
Pulse rate high alarm
limit
Set to 140 BPM
Pulse rate low alarm
limit
Set to 50 BPM
Averaging Time
Set to pre-power down setting
FastSat
Sensitivity**
Display brightness
Pulse tone volume
Alarm Silence