Rad-5/5v Chapter 7: Specifications
www.masimo.com 56 Masimo
Symbols
Definition
Symbols
Definition
Storage Temperature Range
Authorized representative in
the European community
Atmospheric Pressure
Limitation
Keep Dry
Storage/transport relative
humidity range
Fragile/Breakable, Handle with
Care
Instructions/Directions for Use/Manuals are available in electronic format
@http://www.Masimo.com/TechDocs
Note: eIFU is not available in all countries.
Citations
[1] The Masimo rainbow SET technology with Masimo sensors has been validated for no
motion accuracy in human blood studies on healthy adult male and female volunteers with
light to dark skin pigmentation in induced hypoxia studies in the range of 70%-100% SpO
2
against a laboratory CO-Oximeter and ECG monitor.
[2] The Masimo rainbow SET technology with Masimo sensors has been validated for motion
accuracy in human blood studies on healthy adult male and female volunteers with light to
dark skin pigmentation in induced hypoxia studies while performing rubbing and touching
motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5
Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70%-100% SpO
2
against a laboratory CO-Oximeter and ECG monitor.
[3] Masimo SET technology has been validated for low perfusion accuracy in bench top testing
against a Biotek Index 2 simulator and Masimo’s simulator with signal strengths of greater
than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 to 100%.
[4] Masimo rainbow SET technology with Masimo sensors has been validated for pulse rate
accuracy for the range of 25-240 bpm in bench top testing against a Biotek Index 2 simulator.
[5] If the batteries are to be stored for extended periods of time, it is recommended that they
be stored between -20°C to +30°C, and at a relative humidity less than 85%. If stored for a
prolonged period at environmental conditions beyond these limits, overall battery capacity
may be diminished, and lifetime of the batteries may be shortened.
[6] This represents approximate run time at lowest indicator brightness and pulse tone turned
off using new, fully charged batteries. Ensure trending is off to maximize battery life.
[7] With FastSat the averaging time is dependent on the input signal. For the 2 and 4 second
settings the averaging time may range from 2-4 and 4-6 seconds, respectively.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Masimo Rad-5 and is the answer not in the manual?

Masimo Rad-5 Specifications

General IconGeneral
TypeHandheld Pulse Oximeter
SpO2 Range0% to 100%
Pulse Rate Range25 to 240 BPM
Perfusion Index Range0.02% to 20%
Power Source4 AA Alkaline Batteries
Operating Temperature5°C to 40°C
Humidity5% to 95% non-condensing
Storage Temperature-40°C to 70°C (-40°F to 158°F)