Rad-5/5v Chapter 7: Specifications
www.masimo.com 52 Masimo
IEC 61000-3-
3
Guidance and Manufacturer's Declaration - Electromagnetic
Immunity
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The ME Equipment is intended for use in the electromagnetic environment specified
below. The customer or the user of the ME Equipment should assure that it is used in such
an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+6 kV contact
+8 kV air
+6 kV
contact
+8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
---
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s)
to line(s)
± 2 kV line(s)
to earth
---
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT (>95
% dip in UT)
for 0,5 cycle
40 % UT (60
% dip in UT)
for 5 cycles
70 % UT (30
% dip in UT)
for 25 cycles
<5 % UT (>95
% dip in UT)
for 5 s
---
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the [ME
EQUIPMENT or ME SYSTEM] requires
continued operation during power
mains interruptions, it is
recommended that the [ME
EQUIPMENT or ME SYSTEM] be
powered from an uninterruptible
power supply or a battery.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Masimo Rad-5 and is the answer not in the manual?

Masimo Rad-5 Specifications

General IconGeneral
TypeHandheld Pulse Oximeter
SpO2 Range0% to 100%
Pulse Rate Range25 to 240 BPM
Perfusion Index Range0.02% to 20%
Power Source4 AA Alkaline Batteries
Operating Temperature5°C to 40°C
Humidity5% to 95% non-condensing
Storage Temperature-40°C to 70°C (-40°F to 158°F)