Rad-5/5v Chapter 7: Specifications
www.masimo.com 52 Masimo
Guidance and Manufacturer's Declaration - Electromagnetic
Immunity
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The ME Equipment is intended for use in the electromagnetic environment specified
below. The customer or the user of the ME Equipment should assure that it is used in such
an environment.
Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
Mains power quality should be that of
a typical commercial or hospital
environment.
± 1 kV line(s)
to line(s)
± 2 kV line(s)
to earth
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT (>95
% dip in UT)
for 0,5 cycle
40 % UT (60
% dip in UT)
for 5 cycles
70 % UT (30
% dip in UT)
for 25 cycles
<5 % UT (>95
% dip in UT)
for 5 s
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the [ME
EQUIPMENT or ME SYSTEM] requires
continued operation during power
mains interruptions, it is
recommended that the [ME
EQUIPMENT or ME SYSTEM] be
powered from an uninterruptible
power supply or a battery.