Rad-5/5v Chapter 5: Alarms and Messages
www.masimo.com 42 Masimo
Display
Type
Solution
BAD SEN
Defective
sensor
Replace sensor
SEN
blinking
Unrecognized
sensor
Connect appropriate cable
INT DET
Blinking
Interference
detected
Ensure that the sensor is properly applied, and
cover the sensor site with opaque material, if
required.
Rad-5v Messages
The following messages display on Rad-5v:
Display
Message
Type
Solution
NO CBL
No cable connected
Connect cable
INC
CBL
Incompatible cable
Replace cable
RPL
CBL
Defective or unrecognized cable
Replace cable
RPL
CBL
blinking
Cable life expired
Replace cable as soon as possible
NO SEN
No sensor connected
Connect sensor
INC
SEN
Incompatible sensor
Replace sensor

Questions and Answers: