Resistance, Inductance, and Capacitance Test Specifications
Table 22: Resistance test specifications.
Parameter Value
Measurement range 0.001 mΩ–100 kΩ
4-wire measurement Yes
Maximum test current 10 A
Accuracy ±2% ±0.25 mΩ
Table 23: Inductance test specifications.
Parameter Value
Measurement range 0.01 µH–10 H (120 Hz)
0.01µH–200 mH (1000 Hz)
4-wire measurement Yes
Test frequency 120, 1000 Hz
Accuracy ±5% ±5 µH
Table 24: Capacitance test specifications.
Parameter Value
Measurement range 0.01 nF–50 µF
4-wire measurement Yes
Test frequency 4000 Hz
Accuracy ±5% ±1 nF
Table 25: ADX testing and safety standards compliance.
Standard Topic
IEC 61326-1 Ed. 2.0 2012-07 Electrical equipment for measurement, control and laboratory use - EMC
requirements - Table 1
FCC 47CFR: Part 15 Subpart B: 2020 Unintentional Radiators
ICES-003 Issue 7, October 2020 Limits and Methods of Measurement to Information Technology Equipment (including
Digital Apparatus).
IEC 61010-031:2015 Safety requirements for electrical equipment for measurement, control, and
laboratory use. Safety requirements for hand-held probe assemblies for electrical
measurement and test.
IEC 61010-2-034:2017 Safety requirements for electrical equipment for measurement, control, and
laboratory use. Particular requirements for measurement equipment for insulation
resistance and test equipment for electric strength.
IEC 62133-2:2017 Safety Test Standard of Li-Ion
CISPR 11:2009 +A1:2010, Class A Radiated Emissions and AC Mains Conducted Emissions
IEC 61000-3-2:2014 Harmonics
IEC 61000-3-3:2013 Flicker
IEC 61000-4-2:2009 Electro-Static Discharge Immunity Test
IEC 61000-4-3:2010 Radiated, Radio-Frequency, Electromagnetic Immunity
IEC 61000-4-4:2012 Electrical Fast Transient/Burst Immunity Test
IEC 61000-4-5:2006 Immunity to Surges
IEC 61000-4-8:2010 Power Frequency Magnetic Field Immunity Test
IEC 61000-4-11:2004 Voltage Dips/Interruptions Immunity Test
For more safety compliance information, see “EU Declaration of Conformity” in Notices—front section.