58
WIKA operating instructions hydraulic high pressure controller, model CPC8000-H
PN 0018708001B 11/2020 EN
EN
RESET_INSTRU-
MENT_CONTROL_
TIME_ON
Reset the "instrument control time on" counter
to 0.
Resolution <n> Sets the number of signicant digits.
See decpt
Resolution? <sp>n<cr><lf> Returns the number of signicant digits. See
decpt
Rlter Value in % Sets the % of the rate lter.
Rlter? <sp>+n.nnnnnE+nn<cr><lf> Returns the rate lter.
RFilterWin nnn Set the rate lter window as a oating point
value in pressure
RFilterWin? <sp>+n.nnnnnE+nn<cr><lf> Returns the rate lter window
RStable? YES or NO
RStableTime 0 to 65535 Sets the rate stable time to the number of
seconds specied
RStableTime? <sp>XXXXXXX<cr><lf> Returns the rate stable time
RStableWin nnn Alias for “RStableWindow”. See RStableWin-
dow
RStableWin? <sp>+n.nnnnnE+nn<cr><lf> Alias for “RStableWindow?”. See
RStableWindow?
RStableWindow nnn Set the rate stable window as a %FS/second
RStableWindow? <sp>+n.nnnnnE+nn<cr><lf> Returns the rate stable window
Runits Sec, min, hr Sets the rate time unit.
Runits? <sp>XXXX<cr><lf> Returns the rate time unit.
Sbaud 9600, 19200, 38400, 57600,
115200
Sets the serial baud rate.
Sbaud? <sp>XXXX<cr><lf> Returns the serial baud data.
Sdata 7 or 8 Sets the serial data bits.
Sdata? <sp>n<cr><lf> Returns the serial data bits number.
Sensor 1, 2, 3, or PRIMARY,
SECONDARY, TERTIARY
Sets the active transducer.
Sensor? <sp>XXXXXXX,1<cr><lf> Returns active transducer in long string
format.
Sensorid? <sp><sen-
sor#><sp>ID<sp>MEN-
SOR,<sp><mod-
el>,<sp><SN>,<sp>V<ver-
sion><cr><lf>
Returns the active transducer’s serial number
and rmware version.
Setpoint nnn.nnn Sets the control setpoint for the instrument.
Value must be inside upper and lower limits.
Alias for “Setpt”
Setpoint? <sp>nnn.nnnnE+nn<cr><lf> Returns the control setpoint in current units
7. Remote Operation