62
WIKA operating instructions hydraulic high pressure controller, model CPC8000-H
PN 0018708001B 11/2020 EN
EN
Table 7.7
Command or Query Response/Function
MEASure
[:PRESsure] [R]? Returns the pressure from range R.
:TEMPerature[R]? Returns the temperature from range R.
:RATE[R]? Returns the pressure rate/sec from range R.
:BAROmetric? Returns the barometric pressure.
:SLEW[R]? See MEASure:RATE?
CALibration
[:PRESsure][R]
:MODE? Returns 1=calibrated or 0=not calibrated.
:DATE? Returns date of cal “MM/DD/YY”
:DATE <i,i,i> Sets date of cal YYYY,MM,DD
:ZERO? Returns zero oset.
:ZERO <n> Sets the zero oset.
:ZERO:RUN Start autozero sequence.
:ZERO:STOP Stop autozero sequence
:ZERO:INITiate? Returns zero status
:ZERO:INITiate Ignored
SENSe
[:PRESsure] [R]
:NAME? Returns transducer name string.
:MODE? Returns pressure type “ABSOLUTE” or “GAUGE”
:MODE ABS | GAUGE Sets the pressure type to absolute or gauge.
:ABS? Returns native transducer type “1” for absolute, ”0” for gauge.
:RESolution? Returns resolution (oat).
:RANGe
[:UPPer]? Returns the maximum range.
:LOWer? Returns the minimum range.
:Unit
[:NAME]? Returns ASCII units (mixed case).
:VALue? Returns the units conversion factor.
:REFerence
[:HEIGht] <n> Sets the head pressure height.
:HEIGht? Returns the head pressure height.
:MODE? Returns “OFF”, “GAS”, or LIQUID”.
:MODE OFF | GAS | LIQUID Sets the head pressure mode.
:MEDium<n> Sets the medium density.
:MEDium? Returns medium.
7. Remote Operation