WIKA operating instructions hydraulic high pressure controller model CPC8000-H
PN 0018708001B 11/2020 EN
EN
:ACTive <n> Sets the active transducer
:ACTive? Returns the active transducer
SYSTe m
:DATE <i,i,i> Not used, kept for backwards compatibility
:DATE? Not used, does not cause an error, does not return a
response.
:TIME <i,i,i> Not used, kept for backwards compatibility
:TIME? Not used, does not cause an error, does not return a
response.
:ERRor[:NEXT]? Return error code, description.
:KLOCk ON | OFF | 1 | 0 Sets the keylock state.
:PRESet Load default values.
:SAVe no function (not needed).
:VERSion? Return SCPI version 1994.0.
TEST
:ELECtronic? Returns “OK”.
:RELay <n>? Returns status of digital output <n>.
:RELay <n> ON | OFF Turns the digital output on or o <n>.
UNIT ( Note: see Table 11.1 Measurement Units)
:NAME <n>? Returns the units string for units code <n>.
:FACTor <n>? Returns the units conversion for units code <n>.
[:PRESsure] bar|mbar|Pa|psi Sets the pressure units.
[:PRESsure]? Returns the pressure units.
:USER
:FACTOR <n>, <f> Sets the multiplier to <f> for user unit <n>
:FACTOR ? <n> Returns the multiplier <f> for user unit <n>
:BASE <n>, <s> Sets the base engineering units for user unit <n>, options
are psi, bar, pascal.
:BASE? <n> Returns the base engineering units for user unit <n> as a
text string
:INDEX <n> Sets the global remote unit by index number.
:INDEX? Returns the index number.
7. Remote Operation