MI 3000 EasiPLUS Instrument operation
19
Fuse F1 (continuity circuit) blown or not inserted.
S SF F
Single fault condition in IT system.
4.1.4 Result field
Measurement passed.
Measurement failed.
Measurement is aborted. Check the conditions at the
input terminal.
4.1.5 Other messages
Hard Reset
Instrument settings and measurement parameters/limits
are set to initial (factory) values. For more information
refer to chapter 4.5.5. Recalling original settings.
CHECK SUM ERROR
Memory contents damaged. Contact your distributor or
manufacturer for further information.
4.1.6 Sound warnings
The shortest sound
Pressed key deactivated.
Sub-function is not available.
Short sound
Pressed key activated.
Measurement has been started after pressing the TEST
key. Consider any displayed warnings during
measurement.
Long sound
Measurement is prohibited. Consider any displayed
warnings and check online voltage/terminal monitor!
Periodic sound
Warning! Phase voltage on the PE terminal! Stop all the
measurements immediately and eliminate the fault before
proceeding with any activity!