Lighting: selection of 3 levels of back-lighting
of the display unit in order to limit the power
consumption of the multimeter, as follows:
Eco, Normal, Max
The default level of extinction of the
back-lighting is ECO, after 1min if there
has been no action on the front panel of
the multimeter.
An internal accelerometer makes it possible
to wake up the multimeter by simply
touching the product with the adjustment
selected.
Standby: validation (default: yes) or not of
automatic switching off after 30 min, if there
has been no action on the front panel of the
multimeter.
In the SURV, MEM, or Communication
mode, automatic switching off is not
validated.
For your safety, automatic switching off is
disabled when the quantities measured (voltage,
current) on the input exceed the danger
thresholds.
Beep: validation (default) or not of the
emission of an audible signal (beep) when:
- a key is pressed,
- there is a voltage on the "V" input exceeding
605 VDC or 30 VAC,
- a stable measurement is captured in AUTO
HOLD
The audible signal is maintained even
when the buzzer is deactivated:
- in a continuity test,
- when a range is exceeded (voltage or
current),
- for a measurement of 10A or more,
- when there is an incompatibility
between the connections of the leads
and the function selected
- when the supply voltage (battery) is
too low
blinking of the red batt
indicator.
- The audible signal is maintained when
the function is changed while recording
is in progress (low-pitched beep).